Specification of Partech OxyTechw² RDO Pro-X Dissolved Oxygen Sensor Murah
OxyTechw² RDO Pro-X Dissolved Oxygen Sensor
Optical Sensor for Dissolved Oxygen Monitoring
The OxyTechw² RDO Pro-X Sensor uses the latest optical technology for measuring
dissolved oxygen (DO) in demanding wastewater, drinking water, industrial effluent and
surface water environments.
The sensor uses optical sensing techniques that provide a fast response and do not need a
flowing sample to operate. There is no need for routine calibration, enabling manual
intervention to be kept to an absolute minimum. Periodic validation of the measurement
can be done as part of standard quality assurance performed by is the operator. The sensor
uses an extended life cap that does not have a finite life, though replacement is
recommended every two years. When replacement of the cap is required the operation
takes only a few minutes with no need to enter complex set up codes and no need to recalibrate.
The need for manual cleaning of the sensor is kept to a minimum through use of Partech's
FlexTech mounting system which places the sensor below the surface of the tank away from
the floating fats and grease whilst keeping the sensor face moving to prevent the build up of
a biofilm.
The sensor is part of the WaterWatch² family and works with our 7300w² Monitor and can
be combined with our TurbiTechw² LA sensor to provide an extremely cost effective
package for controlling activated sludge plants.
APPLICATIONS
- Activated Sludge Control
- Drinking Water Processing
- Reservoir Water Quality
- River Monitoring
MEASUREMENT PRINCIPAL
- Luminescent Optical Technology
FEATURES
- No drift – no calibration required
- 24 month cap life
- Fast Response
BENEFITS
- Automated Aeration Control
- Low Cost of Ownership
COMPATIBLE MONITOR
MOUNTING OPTIONS
- Flexible Joint for Self Cleaning
ALTERNATIVE SENSORS
PRODUCT WARRANTY
- Probe – 36 months
- Cap – 24 months
Physical
|
Dimensions (hxwxd)
|
Ø47mm, 203 mm long
|
Weight
|
0.8 kg
|
Protection Class
|
IP68
|
Enclosure Material
|
Body - ABS, Titanium Temperature Probe with Polyurethane coated cable
Cap - ABS, Polycarbonate/Polymethyl/methylacrylate
|
Cable Entry
|
Integral Cable Gland
|
Cable Length
|
10 metres standard, 100 metres maximum
|
Interface to Monitor
|
|
Power Supply
|
12 VDC from Monitor
|
Sensor Output
|
ModTechw², RS485
|
Environmental Data
|
|
Operating/Storage Temperature
|
0 to 50ºC
|
Location
|
Indoor/Outdoor
|
Pressure Rating
|
210 mWC
|
Flow
|
No flow necessary
|
Maintenance
|
|
Service
|
The optical cap for the sensor should be replaced every 24 months
Periodic cleaning is required at a frequncy determined by the application. Our self cleaning shaft option is highly recommended, this eliminates manual cleaning between calibrations.
|
Calibration
|
The sensor is inherently stable, calibration should be carried out to suit the application. Our recommendation is yearly
|
Product Publication No. PPPAP/IT00005-18
SKU. 226770 - 226771
Marketing Division
CV. Putra Andalas Prima
Tel. (+62)-61 8836 9476
Fax. (+62)-61 8836 9476
Email. marketing@putraandalasprima.co.id