Specification of Partech TurbiTechw² LR Turbidity Sensor Murah
Partech TurbiTechw² LR Turbidity Sensor
The TurbiTechw2 LR Sensor has been specifically designed to monitor final treated, portable and filter water. The sensor is extremely sensitive and is capable of discriminating changes in Turbidity of less than 0.1 FTU. The 90° light scatter monitoring technique makes the sensor sensitive to a wide range of particle sizes, from colloidal material through to larger particles and raw waters and the large mineral particles sometimes found in distribution samples.
The unique sensor design captures stray light which ensures an ultra stable zero point, this is combined with an automatic self-cleaning mechanism that prevent the build up of fouling within the sample chamber. The measurement system includes a self monitoring loop that compensates for changes in optical performance.
The sensor requires very little maintenance, with no special tools or training required. Calibration can be carried out against either a wet Turbidity standard or using the dry calibration reference cell that is supplied with every sensor. The reference cell provide an extremely easy, repeatable method of checking the performance of the sensor without the cost and inherent errors associated with chemical standards
APPLICATIONS
- Raw Water
- Final Water Monitoring
- Filtered Water
- Filter Backwash Control
- Clarifier Supernatant
- Process Control
MEASUREMENT PRINCIPAL
- 90° Light Scatter
- Wavelength 860nm
- ISO 7027 Compliant
FEATURES
- Dry Secondary Standard
- Ultra Low Back Scatter
- Auto-Clean Sensor
ACCESSORIES
COMPATIBLE MONITOR
SPECIFICATION DATA
Physical
|
Dimensions (hxwxd)
|
440 x 200 x 200 mm
|
Weight
|
4.5 kg
|
Protection Class
|
IP68
|
Enclosure Material
|
Black Acetal Co-Polymer, Polypropylene
|
Seal Material
|
Nitrile
|
Cable Entry
|
Integral Cable Glands
|
Cable Length
|
5 metres standard, 100 metres maximum
|
Process Connections
|
Inlet and Drain: 1/2" BSP with 12 mm hosetail
Outlet: 1/4" BSP with 8 mm hosetail
|
Interface to Monitor
|
|
Power Supply
|
12 VDC from Monitor
|
Sensor Output
|
ModTechw², RS485
|
Environmental Data
|
|
Operating/Storage Temperature
|
0 to 50ºC
|
Location
|
Indoor/Outdoor
|
Pressure Rating
|
2 Bar
|
Maintenance
|
|
Service
|
Wiper and desiccant replacement every year (application dependent)
|
Calibration
|
The sensor is inherently stable, calibration should be carried out to suit the application. Our recommendation is every 6 months.
|
Product Publication No. PPPAP/IT00011-18
Publication Date. 22 March 2018
Marketing Division
CV. Putra Andalas Prima
Tel. (+62)-61 8836 9476
Fax. (+62)-61 8836 9476
Email. marketing@putraandalasprima.co.id