Specification of V 7200 UV VIS NIR Spectrophotometer Murah
The new JASCO V-7200 incorporates an optimized PbS detctor to extend the wavelength range through the NIR to 3300 nm, Users can expect to achieve significant performance breakthroughs for a range of applications from evaluating band pass filters to measuring the high transmission of next generation fiber optic materials.
Specifications:
Wavelength range: 175 - 3300 nm (Low UV kit is necessary for the region below 184 nm)
Bandwidth selectable: 0.049 nm or higher (UV/Vis)
0.2 nm or higher (NIR)
Stray light: 0.00007% or less at 220 nm (10 g/L Nal ASTM Method)
0.00007% or less at 370 nm (50 g/L NaN02)
0.0003% or less at 1420 nm (H20 1 cm path)
0.00048% or less at 2365 nm (CHCl3 1 cm path)